Name/Function

Cardioscope

Manufacture type

Diascope 1

Manufacture

S & W

Made in

Denmark

Part number.

1167400

Production year

1993

Comment's

Singel Channel ECG Monitor.





Contents list: Cardioscope 1167400

Part number

Description

Qty
Note

Contents list

1

Instruction Manual

1

Diascope 1

1

Main cable 220 V

1

Cords for disposable electrodes

1

Cords for multiple use electrodes

Multiple use electrodes

Tape for multiple use electrodes

1

Pack 500 pcs



Technical data

Dimensions

ECG Amplitier

H: 125 x W: 200 x D: 270 mm

Input

Weight

Mode: Differentoal

4.3 kg

Impedance: 10 mOhm DC, 4 mOhm at 35 Hz

Power reguirements

Difference input max: 360 mV

220-240 VAC +/-10% 50/60 Hz or 100-117 VAC +/-

CMRR: 80 db

10% 50/60 Hz

Defi shock protection: 5 KV

Consumption

Filter 50/60 Hz: 40 db

24 VA

Defibrillator marker, input typ: 1 V/100 ms

Operating temperature

Amplification

+5 C to + 40 C

Electrosurgery noise filter

Display

Frequency response: 0.3 – 35 Hz (-3 db)

CRT screen dim. H 80 mm, W: 100 mm.

Overload recovery time 6 s.

Display speed: 25 or 50 mm/s.

Pace discriminator

Max. display gain: 1 mV input equals 2 cm on the screen

Trigger level: +/-2.3 mV, slew rate sensitive

Digital display

Output

3 digit LED

Heart rate test generator: 120 beats/min.

Digit size: H: 11 mm, W 6 mm. Updating cycle: 2.1 s simultaneously

Analogue heart rate: 10mV/beat.

Cathode ray tube

ECG direct (1 mV inp.): 1 Vpp

Size: 5.5” (80 x 100 mm).

ECG delayed (1 mV inp.): 1 Vpp

Mode: Non-fade, rolling mode.

Alarm: =V when active, -11 V when off

Light indications

Alarm block: 0 V when active, -11 V when off 0 V

Alarm off.

Ds 1 on: +5 V when on, -11 V when off.

Controls

Memory back-up

Stop/run for freezing and starting the ECG.

The values set by the operator are retained

Alarm on/off for enabling or disabling the alarm.

During short power-off periods, i.e ECG-gain, alarm

Main function selector for selecting the input of the

on/off, alarm limits

digital display, ECG LOW/HIGH alarm.

The memory time is minimum 15 s.

Up/down button for increasing/decreasing the

Patient safety system

parameters selected by the main functions selector.

Leakage current between chassis and any patient

QRS-volume/ECG trace for adjusting the QRS-beep

terminal at 50 Hz <0.04 uA/V

and the ECG trace speed. Mains on/off.

QRS detector

Soft functions test selector.

Trigger level 0.3 mV (ref. input).

Connectors

Tachometer

ECG input: Colour coded S&W standard connectors

Mode: Averaging.

Interface output: 15 way subminiature D-type

Range: 20-300 beats/min.

Classification

IEC 6001 class 1

Type CF with defibrillator protection

Vibration IEC 68